Acta Mechanica Slovaca 2021, 25(2):14-18 | DOI: 10.21496/ams.2021.019
Displacement Measurement in the Vertical Axis of the Measuring Microscope using Laser Triangulation Sensor
- 1 Technical University of Košice, Faculty of Mechanical Engineering, Department of Biomedical Engineering and Measurement, Letná 1/9, Košice 042 00
The article deals with assessment of the suitability of using a laser triangulation sensor to measure the displacement in vertical axis of a measuring microscope. Data obtained from sensor calibration are used for determination of measuring range in which the smallest measurement error occurs. Linear approximation of the inverse calibration function is applied to correct systematic errors in reduced measuring range of the sensor. Residual measurement errors of corrected sensor output can be applied in uncertainty calculation for sensor future measurements.
Keywords: calibration; laser triangulation sensor; displacement measurement.
Received: March 17, 2021; Revised: April 14, 2021; Accepted: April 21, 2021; Published: June 25, 2021 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Chudý, V., Palenčár, R., Kureková, E., Halaj, M. (1999). Measurement of technical quantities. Slovak University ofTechnology in Bratislava, 1. edition, ISBN 80-227-1275-2
- Gao, S. Dejima, Y. Shimizu, S. Kiyono, H. Yoshikawa (2003). Precision Measurement of Two-Axis Positions and Tilt Motions Using a Surface Encoder, CIRP Annals, Volume 52, Issue 1, Pages 435-438, ISSN 0007-8506
Go to original source...
- H. Kunzmann, T. Pfeifer, J. Flügge: Scales vs. Laser Interferometers Performance and Comparison of Two Measuring Systems (1993). CIRP Annals, Volume 42, Issue 2, Pages 753-767, ISSN 0007-8506
Go to original source...
- Suh, Young. (2019). Laser Sensors for Displacement, Distance and Position. Sensors doi: 10.3390/s19081924.
Go to original source...
- B. Sun, B. Li (2016) Laser Displacement Sensor in the Application of Aero-Engine Blade Measurement, IEEE Sensors Journal, vol. 16, no. 5, pp. 1377-1384
Go to original source...
- Zhifeng Zhang, Qibo Feng, Zhan Gao, Cuifang Kuang, Cheng Fei, Zhang Li, Jinyun Ding (2008) A new laser displacement sensor based on triangulation for gauge real-time measurement, Optics & Laser Technology, Volume 40, Issue 2, 2008, Pages 252-255, ISSN 0030-3992,
Go to original source...
- Kelemenová, T., Dovica, M. (2016). Calibration of measuring instruments. Košice : TUKE, SJF, ISBN: 978-80-553-3069-3
- JCGM 200 - International vocabulary of metrology - Basic and general concepts and associated terms (VIM) 3rd edition (2008 version with minor corrections)
- SNAS: Methodological guideline for accreditation, MSA-L/12, Expression of the uncertainty of measurement in calibration. Bratislava, 2014
This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), which permits use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.