Acta Mechanica Slovaca 2018, 22(1):40-45 | DOI: 10.21496/ams.2018.007
Sensor for Blood Outflow in Intravenous Intervention Detection
- VSB-Technical University of Ostrava, FEECS, K450 17. listopadu 15, 708 33, Ostrava-Poruba, Czech Republic
This paper deals with an issue, resulting in exceptional cases in intraaortic balloon counter pulsation. After analyzing the problem and find a worthy solution for detecting small concentrations of blood mixed with water content in the transparent tube, the first two chapters, optoelectronic detection method chosen. This principle is in the third part, which can be considered as the know-how of the entire thesis, 7 complex experimental measurements verified the accuracy. In this section, selected the best optoelectronic couple, the ideal setting and the algorithm chosen to evaluate the content of the tube. The final chapter is a self-designed device, called INKE1, which is able to repeatedly identify the contents in the tubing from the air, water and up to 0.1% of blood in the water.
Keywords: sensor; blood flow; intravenous intervention detection; counter pulsation.
Published: March 15, 2018 Show citation
References
- Taparia, N., Platten, K.C., Anderson, K.B., Sniadecki, N.J. A microfluidic approach for hemoglobin detection in whole blood (2017) AIP Advances, 7 (10), art. no. 105102.
Go to original source...
- Liou, J.-C., Lin, W.-C., Kong, Y.-Y., Fang, K.-W. Investigated blood flow measurement of lights with different wavelengths system (2017) 2017 IEEE International Conference on Consumer Electronics - Taiwan, ICCE-TW 2017, art. no. 7991014, pp. 99-100.
Go to original source...
- Ramírez-Miquet, E.E., Perchoux, J., Da Costa Moreira, R., Luna-Arriaga, A., Zhaoa, Y., Tronche, C., Sotolongo-Costa, O. Optical feedback interferometry: From basics to applications of laser flowmetry (2017) Revista Cubana de Fisica, 34 (1), pp. 48-57.
- Li, G., Lee, D.-W. An advanced selective liquid-metal plating technique for stretchable biosensor applications (2017) Lab on a Chip, 17 (20), pp. 3415-3421.
Go to original source...
- Luo, C.-H., Su, C.-J., Huang, T.-Y., Chung, C.-Y. Non-invasive holistic health measurements using pulse diagnosis: I. Validation by three-dimensional pulse mapping (2016) European Journal of Integrative Medicine, 8 (6), pp. 921-925.
Go to original source...
- Ullah, H., Ahmad, E., Hussain, F. Optical coherence tomography as glucose sensor in blood (2016) Advanced Structured Materials, 79, pp. 397-428.
Go to original source...
- Fine, I., Kaminsky, A.V., Shenkman, L. A new sensor for stress measurement based on blood flow fluctuations (2016) Progress in Biomedical Optics and Imaging - Proceedings of SPIE, 9707, art. no. 970705.
Go to original source...
- Li, N., Jia, L., Zhang, P. Detection and volume estimation of bubbles in blood circuit of hemodialysis by morphological image processing (2015) Proceedings of the 2015 7th IEEEInternational Conference on Cybernetics and Intelligent Systems, CIS 2015 and Robotics, Automation and Mechatronics, RAM 2015, art. no. 7274625, pp. 228-231.
Go to original source...
- Siegel, G., Meyer-Rath, G., Ermilov, E., Rodríguez, M., Malmsten, M., Claesson, P., Saunders, R., Hetzer, R., Lindman, B. Flow sensing in the cardiovascular system (2015) Colloids and Surfaces A: Physicochemical and Engineering Aspects, 480, pp. 318-327.
Go to original source...
- Aragón, A.M., Hernando-Rydings, M., Hernando, A., Marín, P. Liquid pressure wireless sensor based on magnetostrictive microwires for applications in cardiovascular localized diagnostic (2015) AIP Advances, 5 (8), art. no. 087132.
Go to original source...
This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), which permits use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.